50.病人使用陽壓呼吸器時警報被啟動,呼吸波形如下圖,此警報是屬於那一級? &n..-阿摩線上測驗
1F
|
2F Ollie Hou 高一下 (2019/07/05)
Level 1: Immediately Life-Threatening • Electrical power failuren • No gas delivery to patientn • Exhalation valve failuren • Excessive gas delivery to patientn • Timing failuren Level 2: Potentially Life-Threatening :n • Circuit leakn • Circuit partially obstructedn • Heater/humidifier malfunctionn • I : E ratio inappropriaten • Inappropriate oxygen level (gas/blender failure)n • Autocyclingn • Inappropriate PEEP/CPAP level (too low/too high) Level 3: Not Life-Threatening but a Potential Sourcenof Patient Harm • Changes in lung characteristics (compliance/resistance) • High respiratory rate... 查看完整內容 |
3F
|