69.下列何種裝置不受氣體溫度、黏滯性及密度的影響?
(A)Fleisch pneumotachograph
(B)screen pneumotachograph
(C)vortex ultrasonic flowmeter
(D)variable orifice pneumotachograph
答案:登入後查看
統計: A(36), B(35), C(239), D(49), E(0) #3135296
統計: A(36), B(35), C(239), D(49), E(0) #3135296
詳解 (共 1 筆)
#6092270
1.Fleisch pneumotachograph運用Poiseullie's law,濕氣、氣體黏度會影響,使阻力不固定
2.screen pneumotachograph為 modified Fleisch-type
3.vortex ultrasonic flowmeter運用超音波被氣流干擾程度來推算flow
4.variable orifice pneumotachograph運用孔洞(創造turbulent flow)兩端壓差來換算flow
〈補充〉Turbinometer利用氣流轉動turbine,以此產生的動能換算flow。因此不受擾流、氣體組成、溫度、水蒸氣影響測量值
3