33.Puritan Bennett 7200ae於PCV mode,其設定值為:壓力:20 cm H2 O;Ti = constant 1.0 second;rate =10 bpm;mode = CMV;若將rate降到6 bpm,病人未啟動呼吸器的狀況下,I:E ratio會有何變化?
(A)1:5
(B)1:6
(C)1:9
(D)機器不會有任何顯示
詳解 (共 1 筆)
未解鎖
TCT=(60/6)=1010(TCT)...