36.一病人接受腹部電腦斷層檢查,偵檢器組置(collimation)為64 × 0.5 mm,X光旋轉一圈為0.5 s,檢查床移 動速度為25.6 mm/s,管電流為200 mA,管電壓為120 kVp,影像重組厚度為5 mm,有效管電流時間乘積 (effective mAs)為何?
(A)80 mAs
(B)100 mAs
(C)125 mAs
(D)250 mAs
答案:登入後查看
統計: A(240), B(693), C(624), D(713), E(0) #1635382
統計: A(240), B(693), C(624), D(713), E(0) #1635382
詳解 (共 5 筆)
#2598953
200x0.5/[(25.6x0.5)/(64x0.5)]=250